:VOLTage[:DC]
CONFigure:VOLTage[:DC] [<range>[,<resolution>]] [,<channel_list>] selects the
DC voltage function and allows you to specify the range and resolution. If you
specify a channel list, those multiplexer channels are scanned.
For a complete listing of range and resolution values available, see
Table 5-1 on page 120.
Parameters
Parameter Name Parameter Type Range of Values Default Units
<range>
numeric 0.113 V | 0.91 V | 7.27 V |
58.1 V | 300 V | AUTO | DEF |
MIN | MAX
volts
<resolution>
numeric
resolution (see Table 5-1) |
DEF | MIN | MAX
volts
<channel_list>
numeric Refer to the multiplexer user’s manual for
a list of channels available.
none
Comments The :DC parameter is optional. Both of the following command statements select
the DC voltage function:
CONF:VOLT:DC
or CONF:VOLT
To select a standard measurement range, specify range as the input signal’s
maximum expected voltage. The multimeter then selects the correct range.
The AUTO and DEFault options for the range parameter have the same effect
(enable autorange). The
DEF option for the resolution parameter defaults the
integration time to 1 PLC.
The MIN and MAX parameters select the minimum or maximum values for range
and resolution:
For range:
MIN = 0.113 V; MAX = 300 V.
For resolution:
MIN selects the best resolution (the smallest value from Table 5-1) for
the selected range.
MAX selects the worst resolution (the largest value from Table 5-1)
for the selected range.
The channel list is of the form (@ccnn), (@ccnn,ccnn), or (@ccnn:ccnn), where
cc = card number and nn = channel number (105 is channel 05 of card number 1,
for example).
To select autorange, specify AUTO (or DEFault) for range or do not specify a
value for the parameter. In the autorange mode, the multimeter samples the input
signal before each measurement and selects the appropriate range.
To specify a MIN or MAX resolution while autoranging, you must specify
CONF:VOLT:DC AUTO or CONF:VOLT:DC DEF (you cannot omit the range
Chapter 5 Command Reference 132