parameter). This prevents the MIN or MAX resolution from being interpreted as a
range setting.
The fastest aperture time available when autoranging is 100 µs. In order to specify an
aperture time of 10 µs, you must select a fixed range.
Related Commands: FETCh?, INITiate, READ?
Example Making DC Voltage Measurements
CONF:VOLT 0.825,MAX,(@100:103) !Function: DC voltage;
range selected: 0.91 V;
MAX resolution: 61.035
µ
V; specify
channel list.
TRIG:COUN 3 !Scan channel list 3 times
(take 4 readings per trigger).
INIT !Place multimeter in wait-for-trigger state;
store readings in mainframe memory; trigger
source is IMMediate by default.
FETC? !Place readings in output buffer.
enter statement !Enter readings into computer.
133 Command Reference Chapter 5