User`s manual
DISPlay
The DISPlay command subsystem monitors the state of the selected multiplexer
channel within the scanning multimeter. This command is useful only with
mainframes that have a front panel display, such as the HP 75000 Series B
Mainframe (Model HP E1301A).
Subsystem Syntax DISPlay
:MONitor
:CHANnel <channel>
:CHANnel?
[:STATe] <mode>
[:STATe]?
:MONitor:CHANnel
DISPlay:MONitor:CHANnel <channel> selects a single multiplexer channel to be
monitored. Use the
DISPlay:MONitor:STATe command to enable and disable the
monitor mode.
Parameters
Parameter Name Parameter Type Range of Values Default Units
<channel>
numeric Refer to the multiplexer user’s manual for
a list of channels available.
none
Comments • Use the channel parameter to specify a single multiplexer channel within the
scanning multimeter. The channel list is of the form
(@ccnn), where cc = card
number and
nn = channel number
(105 is channel 05 of card number 1, for example).
• Use AUTO in place of channel to display measurements from the most recent
channel to receive a
CONFigure or MEASure command. The channel number and
measurement are updated as the scan progresses. You may want to add a small
delay to the scan or use a slow, externally paced scan in order to view each
channel from the mainframe’s front panel.
• The following example shows the multimeter’s monitor mode display on the HP
E1301A front panel.
• Related Commands: DISPlay:MONitor:STATe
•
*RST Condition: DISP:MON:CHAN AUTO
VOLTMTR_8: Chan: 101 +1.465302E-01
mon volt
Chapter 5 Command Reference 136