Comments The [:DC] parameter is optional. Both of the following command statements
select the DC voltage function:
MEAS:VOLT:DC? or MEAS:VOLT?
To select a standard measurement range, specify range as the input signal’s
maximum expected voltage. The multimeter then selects the correct range.
The AUTO and DEFault options for the range parameter have the same effect
(enable autorange). The
DEF option for the resolution parameter defaults the
integration time to 1 PLC.
The MIN and MAX parameters select the minimum or maximum values for range
and resolution:
For range:
MIN = 0.113 V; MAX = 300 V.
For resolution:
MIN selects the best resolution (the smallest value from Table 5-1)
for the selected range.
MAX selects the worst resolution (the largest value from
Table 5-1) for the selected range.
The channel list is of the form (@ccnn), (@ccnn,ccnn), or (@ccnn:ccnn), where
cc = card number and nn = channel number (105 is channel 05 of card number 1,
for example).
To select autorange, specify AUTO (or DEF) for range or do not specify a value
for the parameter. In the autorange mode, the multimeter samples the input signal
before each measurement and selects the appropriate range.
To specify a MIN or MAX resolution while autoranging, you must specify
MEAS:VOLT:DC? AUTO or MEAS:VOLT:DC? DEF (you cannot omit the range
parameter). This prevents the
MIN or MAX resolution from being interpreted as a
range setting.
The fastest aperture time available when autoranging is 100 µs. In order to specify an
aperture time of 10 µs, you must select a fixed range.
Example Making DC Voltage Measurements
MEAS:VOLT:DC? 0.825,MAX,(@100) !Function: DC voltage;
range selected: 0.91 V;
MAX resolution: 61.035 µV; specify single
channel; trigger source is IMMediate by
default.
enter statement !Enter reading into computer.
149 Command Reference Chapter 5