Related Commands: SAMPle:COUNt, SAMPle:SOURce, SENSe
*RST Condition: SAMPle:TIMer 50E-3 seconds
Example Setting the Sample Period
CONF:VOLT:DC !Function: DC voltage; stand-alone
multimeter.
SAMP:COUN 10 !Specify 10 readings per trigger.
SAMP:SOUR TIM !Sample source is SAMPle:TIMer command.
SAMP:TIM 0.065 !Set 65 ms sample period.
READ? !Place multimeter in wait-for-trigger state and
make measurements; send readings to output
buffer.
enter statement !Enter readings into computer.
:TIMer?
SAMPle:TIMer? [MIN | MAX] returns one of the following numbers to the output
buffer:
The present sample period (76 µs through 65.534 ms) if MINimum or MAXimum is
not specified.
The minimum sample period available (76 µs) if
MIN is specified.
The maximum sample period available (65.534 ms) if
MAX is specified.
Example Querying the Sample Period
SAMP:SOUR TIM !Sample source is SAMPle:TIMer command.
SAMP:TIM MAX !Set sample period to maximum.
SAMP:TIM? !Query multimeter to return sample period.
enter statement !Enter value into computer.