User`s manual

The greater the number of PLCs, the greater the normal mode rejection (and the
lower the reading rate).
Only the 1 PLC and 16 PLC settings provide normal mode rejection of 50 Hz or
60 Hz power line related noise.
*RST Condition: 1 PLC
Example Setting the Integration Time in PLCs
VOLT:NPLC 16 !Integration time is 16 PLCs.
VOLTage:NPLC?
[SENSe:]VOLTage:NPLC? [MIN | MAX] returns one of the following numbers to
the output buffer:
The present integration time in PLCs if
MIN or MAX is not specified.
The minimum integration time available (0.0005) if
MIN is specified.
The maximum integration time available (16) if
MAX is specified.
Example Querying the Integration Time
VOLT:NPLC 16 !Integration time is 16 PLCs.
VOLT:NPLC? !Query multimeter to return integration time.
enter statement !Enter value into computer.
VOLTage:RANGe:AUTO
[SENSe:]VOLTage:RANGe:AUTO <mode> enables or disables the autorange
function for voltage measurements.
Parameters
Parameter Name Parameter Type Range of Values Default Units
<mode>
boolean OFF | 0 | ON | 1 none
Comments You can substitute decimal values for the OFF ("0") and ON ("1") parameters.
When autoranging is ON, the multimeter samples the input before each
measurement and selects the appropriate range.
If you explicitly select a range using VOLTage:AC:RANGe or
VOLTage:DC:RANGe, autoranging is turned OFF.
In order to specify an aperture time of 10 µs, you must select a fixed range
(
VOLT:RANG:AUTO OFF, for example).
Chapter 5 Command Reference 174