194 1326B/E1411B Multimeter Specifications Appendix A
True RMS AC Voltage (AC coupled)
Crest Factor: 7 at 10% full scale; 1.5 at full scale
DC to 60 Hz common mode rejection (CMR)
64 V and 300 V range 50 dB
All other ranges 86 dB
DC to 400 Hz common mode rejection (CMR)
64 V and 300 V range 34 dB
All other ranges 70 dB
Conditions: CMR and Effective Common Mode Rejection (ECMR) measured with 1 k in each of HIGH and LOW leads, 10%
imbalance. LOW connected to COMMON at source. Measured with respect to earth ground.
Resolution vs. Aperture (volts)
Range Input Impedance 267/320 ms 16.7/20 ms 2.5 ms 100 µs 10 µs
87.5 mV >100 mΩ, <100pF 0.03 µV 0.12 µV 0.5 µV 4 µV 7.6 µV
700 mV >100 mΩ, <100pF 0.24 µV 1 µV 4 µV 30 µV 61 µV
5.6 V >100 mΩ, <100pF 2 µV 7.6 µV 30 µV 250 µV 488 µV
44.8 V 10 M±5%, <100pF 15 µV 61 µV 250µV 2 mV 3.9 mV
300 V 10 M±5%, <100pF 122 µV 488 µV 2 mV 16 mV 31 mV
Accuracy conditions for table below: Autozero on, one-hour warmup. Within 24 hours and ±1°C of
calibration temperature (module calibrated at 18°C to 28°C). Sine wave inputs >10% of full scale; DC
component <10% of AC component.
24-Hour Accuracy vs. Aperture ± (% of reading + volts)
Range Frequency 267/320 ms All other
apertures
87.5 mV 20 - 50 Hz
50 Hz - 1 kHz
1 - 5 kHz
5 - 10 kHz
2% + 200 µV
.5% + 200 µV
.5% + 200 µV
3% + 200 µV
N/A
.5% + 200 µV
.5% + 200 µV
3% + 200 µV
700 mV 20 - 50 Hz
50 Hz - 1 kHz
1 - 5 kHz
5 - 10 kHz
2% + 1.5 mV
.5% + 1.5 mV
.5% + 1.5 mV
3% + 1.5 mV
N/A
.5% + 1.5 mV
.5% + 1.5 mV
3% + 1.5 mV
5.6 V 20 - 50 Hz
50 Hz - 1 kHz
1 - 5 kHz
5 - 10 kHz
2% + 15 mV
.5% + 15 mV
.5% + 15 mV
3% + 15 mV
N/A
.5% + 15 mV
.5% + 15 mV
3% + 15 mV
44.8 V 20 - 50 Hz
50 Hz - 1 kHz
1 - 5 kHz
5 - 10 kHz
2% + 100 mV
.5% + 100 mV
1% + 100 mV
10% + 100 mV
N/A
.5% + 100 mV
1% + 100 mV
10% + 100 mV
300 V 20 - 50 Hz
50 Hz - 1 kHz
1 - 5 kHz
5 - 10 kHz
2% + 500 mV
.5% + 500 mV
1% + 500 mV
10% + 500 mV
N/A
.5% + 500 mV
1% + 500 mV
10% + 500 mV