User`s manual

Appendix B
HP E1326B/E1411B Multimeter Error
Messages
The error messages associated with the HP E1326B/HP E1411B multimeter
are shown in this appendix.
Code Message Cause
- 101 Invalid character Unrecognized character in specified parameter.
- 102 Syntax error Command is missing a space or comma between parameters.
- 103 Invalid separator Command parameter is separated by a space rather than a comma.
- 104 Data type error The wrong data type (i.e., number, character, string, expression) was
used when specifying a parameter.
- 108 Parameter not allowed Parameter specified in a command which has only a command header.
- 109 Missing parameter No parameter specified in the command in which a parameter is
required.
- 113 Undefined header Command header was incorrectly specified.
- 124 Too many digits 257 digits were specified for a parameter.
- 128 Numeric data not allowed A number was specified for a parameter when a letter is required.
- 131 Invalid suffix Parameter suffix incorrectly specified (e.g., 5 K rather than 5 KOHM).
- 138 Suffix not allowed Parameter suffix is specified when one is not allowed.
- 141 Invalid character data The parameter type specified is not allowed
(e.g., MEAS:TEMP? TC,O - O is not a choice).
- 178 Expression data not allowed A parameter other than the channel list is enclosed in parentheses.
- 211 Trigger ignored Trigger occurred while the multimeter is in the idle state, or a trigger
occurred from a source other than the specified source.
- 213 INIT ignored An INIT command is received when the multimeter is already in the
wait-for-trigger state following TRIG:SOUR HOLD.
- 214 Trigger deadlock The multimeter is triggered from another source (e.g., READ?) after
the trigger source has been set to TRIG:SOUR BUS.
- 221 Settings conflict Multimeter parameters are set such that a measurement cannot be
made (e.g., specifying a fixed resolution while autoranging).
- 222 Data out of range The parameter value specified is too large or too small.
- 224 Illegal parameter value The numeric value specified is not allowed (e.g., MEAS:TEMP? RTD).
Appendix B HP E1326B/E1411B Multimeter Error Messages 197