Code Message Cause
- 230 Data corrupt or stale Data in mainframe memory is fetched after a command
(e.g., MEASure, READ?) has returned data to the
output buffer.
- 231 Data questionable Resolution is too great for specified range.
Measurement is still taken.
- 240 Hardware error Hardware error detected during power-on cycle. Return
multimeter to Hewlett-Packard for repair.
- 270 Macro error Deleting a macro that does not exist.
- 273 Illegal macro label Macro name begins with a character other than a letter.
- 276 Macro recursion error The macro called is nested too deep within other macro
calls.
- 277 Macro redefinition not allowed A macro with that label already exists.
- 300 Device specific error Severe overload, protection relay is opening (caused
by fixed range and severe overload or COM not
connected to Lo). To fix, use autorange, higher range,
or connect COM to Lo.
- 310 System error Out of macro memory.
- 350 Too many errors The error queue is full as more than 30 errors have
occurred.
- 410 Query interrupted Data is not read from the output buffer before another
command is executed.
- 420 Query unterminated Command which generates data not able to finish
executing due to a multimeter configuration error.
- 430 Query deadlocked Command execution cannot continue since the
mainframe’s command input
+ 1000 Out of memory Not enough memory to store the number of
measurements requested.
+ 2000 Invalid card number There is no multiplexer which corresponds to the card
number specified in the channel list.
+ 2001 Invalid channel number The channel or range of channels specified for the
(@channel list) parameter does not exist.
+ 2008 Scan list not initialized Occurs when INIT is executed after error 2000 or 2001
occurred.
+ 2009 Too many channels in channel list The number of channels in the specified channel list is
larger than the number of channels on the
multiplexer(s).
+ 2012 Invalid channel range Specifying a channel range such as (@115:100).
+ 2600 Function not supported on this card Function specified cannot be measured (specifying a
2-wire resistance measurement without a multiplexer,
for example).
+ 2601 Channel list required Measurement function specified requires a channel list
(specifying a thermocouple measurement without a
multiplexer, for example).
+ 2602 Timer too fast Aperture time is longer than sample rate.
198 HP E1326B/E1411B Multimeter Error Messages Appendix B