380 OUTPUT 70903;"INIT"
390 !
Retrieve the readings from multimeter memory and enter them into the
400 !controller. Clear the switchbox to exit the continuous scanning mode.
410 OUTPUT 70903;"FETC?"
420 ENTER 70903;Rdgs(*)
430 CLEAR 70914
440 END
Comments
The multimeter at secondary address 03 (logical address 24) is
connected to the switchbox at secondary address 14 (logical
address 112) with an analog bus cable and a digital bus cable.
The analog bus carries the input signals to the multimeter. The
digital bus is used to carry a "multimeter complete" signal to the
switchbox to trigger the next channel closing.
Because of the fast rate at which the FET channels close, the
multimeter is triggered once (
INIT) and then samples continuously
(
SAMP:COUN 500). Thus, the multimeter ignores the multiplexer
"channel closed" signal on the digital bus.
Using this configuration, the multimeter is able to continuously scan the
switchbox and store readings in its memory at a 76 µs (13 kHz) rate.
Chapter 3 Using the HP E1326B/E1411B Multimeter 51