User`s manual

The terms MIN and MAX often appear as parameter choices in a
command’s syntax.
MIN selects the minimum numeric value for that
parameter.
MAX selects the maximum numeric value. This
eliminates the need to look up specific numbers in the manual.
In this program, note that MAX in the CONFigure command selects
the least resolution and sets the aperture time to 10 µs (see Table 4-5
on page 92). When measurement speed is critical, readings should
be stored in mainframe memory first, rather than returned directly to
the output buffer.
The total number of readings which can be stored depends on the
amount of memory available. Each reading stored will consume four
bytes of memory.
To increase the (throughput) speed at which measurement data is
transferred from the multimeter to the computer by the
FETCh?
command, the multimeter’s output data format should be set to
REAL,64 or REAL,32 (see “Changing the Data Format” on page 58).
The 13 kHz reading rate must be reduced to 12.82 kHz when the
sample count is greater than 32 k. Setting the
SAMP:TIM to 78 µs
gives a reading rate of 12.82 kHz and allows the sample count to be
greater than 32 k.
CONF:VOLT:DC 7.27,MAX
CAL:ZERO:AUTO OFF
SAMP:SOUR TIM
SAMP:TIM 0.078
INIT
Chapter 3 Using the HP E1326B/E1411B Multimeter 57