User`s manual

Four-wire measurements are specified as:
TEMP FRTD,type
The function appears in the
MEASure and CONFigure commands as:
MEAS:TEMP? FRTD,type[,(@channel_list)]
CONF:TEMP FRTD,
type[,(@channel_list)]
Note When making temperature measurements with the MEASure command, the
question mark (?) must be inserted between
TEMP and the temperature
transducer used. Also, if a channel list immediately follows the transducer, it
must be separated by a comma (,) (e.g.
MEAS:TEMP? THER,5000,(@100)).
Specifying a
Function
The measurement functions described previously are represented by the
"measurement" parameter in the
MEASure and CONFigure commands:
MEASure:measurement?
[range|AUTO|DEF|MIN|MAX[,resolution|DEF|MIN|MAX]][,(@channel_list)]
CONFigure:
measurement
[range|AUTO|DEF|MIN|MAX[,resolution|DEF|MIN|MAX]][,(@channel_list)]
When using the stand-alone multimeter, the low-level command:
FUNCtion:function
can be used to change the measurement function without causing a complete
reconfiguration of the multimeter. The stand-alone multimeter functions
which can be changed are:
VOLT:DC
VOLT:AC
FRESistance
The next example shows you how to change from a DC voltage
measurement to a 4-wire resistance measurement.
90 Understanding the HP E1326B/E1411B Multimeter Chapter 4