160 PAUSE 
170 CLEAR SCREEN 
180 ! 
190   
! Measure DC Isolation (HI to Chassis, HI to LO, HI to GU) 
200 ! 
210 OUTPUT @Mux;"*RST" 
220 FOR I=0 TO 2 
230  PRINT TABXY(1,1),"DC Isolation HI to ";Conn$(I);" Measurements
" 
240  PRINT TABXY(1,3),"1. Connect DMM INPUT HI lead to COMMON
HI" 
250  PRINT TABXY(1,4),"2. Connect DMM INPUT LO to ";Conn$(I+3) 
260  DISP " Press Continue when connections are complete " 
270  PAUSE 
280  CLEAR SCREEN 
290  OUTPUT @Mux;"CLOS (@100:115,190:193)" 
300  OUTPUT @Dmm;"TRIG SGL" 
310  ENTER @Dmm;Value(I) 
320 NEXT I 
330 DISP " Press Continue to display measurement results " 
340 PAUSE 
350 CLEAR SCREEN 
360 PRINT TABXY(1,1),"DC Isolation Tests "
370 PRINT TABXY(1,3),"HI to CHASSIS (Ohms)";Value(0) 
380 PRINT TABXY(1,4),"HI to LO (Ohms)  ";Value(1)
390 PRINT TABXY(1,5),"HI to GUARD (Ohms) ";Value(2) 
400 END
Typical Result  A typical result for an overload on all three measurements is :
DC Isolation Tests
HI to Chassis(Ohms)   1.E+38
HI to LO (Ohms)       1.E+38
HI to GUARD (Ohms)   1.E+38 
36 Verification Tests Chapter 2