User`s manual
Chapter 3
Using the Agilent E1346A Multiplexer
Module
Using This Chapter
This chapter uses typical examples to show how to use the 48-Channel
Relay Multiplexer Module. Refer to Chapter 4 (Understanding the
48-Channel Multiplexer Module) for more information.  This chapter
contains the following sections:
• Multiplexer Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  Page 25
• Connecting Channels to Common for Making
Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  Page 26
• Scanning a Range of Switchbox Channels . . . . . . . . . . . . . .  Page 29
Multiplexer Commands
    Table 3-1. Multiplexer Commands in Chapter 3 
Command 
Description
ARM:COUN<number>  Selects multiple scanning cycles
INIT[:IMM]  Starts scanning process; use the [ROUT:]SCAN command
to select channels for scanning
INIT:CONT 1 |0 |ON |OFF  Enables/disables continuous scanning cycles
[ROUT:]CLOS<channel_list>  Closes all channels in channel list
[ROUT:]CLOS?<channel_list>  Queries mainframe for channel closure
[ROUT:]OPEN<channel_list>  Opens all channels in channel list
[ROUT:]SCAN<channel_list>  Closes all channels in channel list one at a time; previous
channel opens before next channel closes
[ROUT:]SCAN:MODE RES  Automatically closes the DT Tree Switch for 2-wire ohms
measurements while scanning
NOTE: The commands with brackets ([ ]) are implied and are not shown in this chapter
Chapter 3 Using the Agilent E1346A Multiplexer Module 25