Service manual

LO to Chassis Leakage  1. Make Hardware Connections
• Turn power supply and mainframe power OFF
• Connect DMM, power supply, and resistor as shown in 
Figure 2-10
• Turn power supply and mainframe power ON
2. Check Direct Terminals Leakage Current
• Send *RST to FET Multiplexer
• Send TRIG SGL to DMM
• Record the DMM reading in Table 2-1 
Positive Polarity, LO to Chassis, Direct
• The DMM measurement should be less than 0.010 Vdc. 
A measurement out of this range indicates a failure of the 
FET Multiplexer. 
3. Check Channels Leakage 
• Send CLOS(@nn00) to the FET Multiplexer, where nn is the
card number (typically 01)
• Send TRIG SGL to DMM
Agilent E1300B/E1301B
Figure 2-10. Positive LO to Chassis Leakage Connections
Verification Tests 2-17