Agilent 75000 Series B Agilent E1364A 16-Channel Form C Switch Module Service Manual Serial Numbers This manual applies directly to instruments with serial numbers prefixed with 2934A and higher. Copyright © Agilent Technologies, Inc. 1991-2005 *E1364-90012* E1364-90012 E012 Manual Part Number: Ε1364−90012 Printed September 2012 Edition 0912 Printed in U.S.A.
Contents Agilent E1364A Switch Module Service Manual Warranty . . . . . . . . . . WARNINGS . . . . . . . . Safety Symbols . . . . . . Declaration of Conformity . Reader Comment Sheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Chapter 4. Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Introduction . . . . . . . . . . . . . Equipment Required . . . . . . . Service Aids . . . . . . . . . . . Troubleshooting Techniques . . . . Identifying the Problem . . . . . Testing the Assembly . . . . . . Repair and Maintenance Guidelines . ESD Precautions . . . . . . . . . Soldering Printed Circuit Boards . Post-Repair Safety Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Certification Agilent Technologies, Inc. certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (formerly National Bureau of Standards), to the extent allowed by that organization’s calibration facility, and to the calibration facilities of other International Standards Organization members.
Documentation History All Editions and Updates of this manual and their creation date are listed below. The first Edition of the manual is Edition 1. The Edition number increments by 1 whenever the manual is revised. Updates, which are issued between Editions, contain replacement pages to correct or add additional information to the current Edition of the manual. Whenever a new Edition is created, it will contain all of the Update information for the previous Edition.
Declaration of Conformity Declarations of Conformity for this product and for other Agilent products may be downloaded from the Internet. There are two methods to obtain the Declaration of Conformity: • Go to http://regulations.corporate.agilent.com/DoC/search.htm . You can then search by product number to find the latest Declaration of Conformity. • Alternately, you can go to the product web page (www.agilent.
Chapter 1 General Information Introduction This manual contains information required to test, troubleshoot, and repair the Agilent E1364A B-Size VXI Form C Switch. See the Agilent E1364A User’s Manual for additional information. Figure 1-1 shows the E1364A Switch Module. Figure 1-1.
Safety Considerations This product is a Safety Class I instrument that is provided with a protective earth terminal when installed in the mainframe. The mainframe, switch, and all related documentation should be reviewed for familiarization with safety markings and instructions before operation or service. Refer to the WARNINGS on page 6 in this manual for a summary of safety information. Safety information for preventive maintenance, testing, and service follows and is also found throughout this manual.
WARNING USING AUTOTRANSFORMERS. If the mainframe is to be energized via an autotransformer (for voltage reduction) make sure the common terminal is connected to neutral (that is, the grounded side of the main’s supply). CAPACITOR VOLTAGES. Capacitors inside the mainframe may remain charged even when the mainframe has been disconnected from its source of supply. USE PROPER FUSES.
Inspection/Shipping This section contains initial (incoming) inspection and shipping guidelines for the E1364A Switch Module. Initial Inspection WARNING Use the steps in Figure 1-2 as guidelines to perform initial inspection of the switch module. To avoid possible hazardous electrical shock, do not perform electrical tests if there are signs of shipping damage to the shipping container or to the instrument. Figure 1-2.
Shipping Guidelines Follow the steps in Figure 1-3 to return the E1364A Switch Module to an Agilent Technologies Sales and Support Office or Service Center. 1. Prepare the Switch • Remove user wiring from terminal module. • Attach tag to switch that identifies: – Owner – Model Number/Serial Number – Service Required • Place tagged device in approved anti-static bag. 2. Package the Switch • Place packaged switch in shipping carton.
Environment The recommended operating environment for the E1364A Switch Module is: Environment Temperature Humidity Operating 0oC to +55oC <65% relative (0oC to +40oC) Storage and Shipment -40oC to +75oC <65% relative (0oC to +40oC) Switch Description The Agilent E1364A Switch Module is an "instrument" in the slots of a VXIbus mainframe. As such, it is assigned an error queue, input and output buffers, and a status register. The switch has 16 channels of Form C relays.
Recommended Test Equipment Table 1-1 lists the test equipment recommended for testing, adjusting, and servicing the E1364A Switch Module. Essential requirements for each piece of test equipment are described in the "Requirements" column. Table 1-1. Recommended Test Equipment Instrument Requirements Recommended Model Use* Controller, GP-IB GP-IB compatibility as defined by IEEE Standard 488-1978 and the identical ANSI Standard MC1.
Notes 18 General Information Chapter 1
Chapter 2 Verification Tests Introduction The three levels of test procedures described in this chapter are used to verify that the E1364A Switch Module: • is fully functional (Functional Verification) • meets selected testable specifications (Operation Verification) • meets all testable specifications (Performance Verification) Test Conditions and Procedures See Table 1-1 for test equipment requirements. You should complete the Performance Verification tests at least once a year.
Functional Verification The Functional Verification Test for the E1364A switch consists of sending the *IDN? command and checking the response. This test can be used at any time to verify that the switch is connected properly and is responding to basic commands. Procedure 1. Verify that the switch is installed in the mainframe and that the mainframe has passed its power-on test. 2. Send the *IDN? command to the switch module (see example following). 3.
Performance Verification The procedures in this section are used to test the switch module’s electrical performance using the specifications in Appendix A of the E1364A User’s Manual as the performance standards. These tests are suitable for incoming inspection, troubleshooting, and preventive maintenance. Test Fixture A test fixture is required for the following tests.
Figure 2-1(b).
Test 2-1: Closed-Channel Resistance Test The purpose of this test is to verify that all relay contacts meet the closed-channel resistance specification for the switch module. If the closed-channel resistance of any contact is greater than 3.5 Ω, the relay should be replaced. Normally Open Contacts 1. Hardware Connections Figure 2-2. Closed-Channel Resistance (NO channels) 2. Equipment Setup • Set DMM to: 4-wire ohms, autorange. • Send *RST to the switch to open all NO contacts. 3.
Normally Closed Contacts 1. Hardware Connections Figure 2-3. Closed-Channel Resistance (NC Channels) 2. Equipment Setup • Set DMM to: 4-wire ohms, autorange. • Send *RST to the switch. • Send CLOS (@100:115) to the switch to open all NC contacts. 3. Closed-Channel Reading (channel 00) • Send OPEN (@100) to the switch to close channel 00. • Trigger the DMM and record the reading. 4. Open-Channel Reading (channel 00) • Send CLOS (@100) to the switch to open channel 00.
Example: Closed-Channel Resistance Test This example performs the Closed-Channel Resistance Test for all Normally Open and Normally Closed contacts.
380 390 400 410 420 430 440 450 460 470 480 490 500 510 520 530 540 26 Verification Tests OUTPUT @Switch;"CLOS (@100:115)" !Open all NC contacts FOR I=0 TO 15 !Closed-channel resistance (NC contacts) OUTPUT @Switch;"OPEN (@"&VAL$(100+I)&")" OUTPUT @Dmm;"TRIG SGL" ENTER @Dmm;Rdg PRINT "Channel "&VAL$(I)&" (closed): "&VAL$(Rdg) ! !Open-channel resistance (NC contacts) OUTPUT @Switch;"CLOS (@"&VAL$(100+I)&")" OUTPUT @Dmm;"TRIG SGL" ENTER @Dmm;Rdg PRINT "Channel "&VAL$(I)&" (open): "&VAL$(Rdg) PRINT NEXT I !
Test 2-2: DC Isolation Test This test verifies that sufficient DC isolation exists between various points on the switch module. The DMM used should be capable of measuring up to at least 1 GΩ. If the DMM indicates an overload, record the reading as ">Rmax", where Rmax is the highest resistance that the DMM can measure. If the DMM is a 3458A, for example, the reading should be written as ">1.2 GΩ". Normally Open to Common 1. Hardware Connections Figure 2-4. DC Isolation Test (NO Channels to Common) 2.
Normally Closed to Common 1. Hardware Connections Figure 2-5. DC Isolation Test (NC Channels to Common) 2. Equipment Setup • Set DMM to: 2-wire ohms, 1 GΩ range. • Send *RST to the switch. • Send the CLOS (@100:115) command to the switch to open all NC contacts. 3. DC Isolation Reading • Trigger the DMM and record the reading.
Normally Open to Chassis 1. Hardware Connections Figure 2-6. DC Isolation Test (NO Channels to Chassis) 2. Equipment Setup • Set DMM to: 2-wire ohms, 1 GΩ range. • Send *RST to the switch. • Send the CLOS (@100:115) command to the switch to close all NO contacts. 3. DC Isolation Reading • Trigger the DMM and record the reading.
Normally Closed to Chassis 1. Hardware Connections Figure 2-7. DC Isolation Test (NC Channels to Chassis) 2. Equipment Setup • Set DMM to: 2-wire ohms, 1 GΩ range. • Send *RST to the switch to close all NC contacts. 3. DC Isolation Reading • Trigger the DMM and record the reading.
Example: DC Isolation Test This example performs the DC Isolation Test for Normally Open and Normally Closed channels to both Common and chassis (earth ground).
410 OUTPUT @Dmm;"TRIG SGL" 420 ENTER @Dmm;Rdg 430 ! 440 PRINT "DC Isolation -- Normally Open Channels to Chassis" 450 PRINT "R = "&VAL$(Rdg) 460 PRINT 470 ! 480 !---------------NORMALLY CLOSED TO CHASSIS--------------490 DISP "Connect DMM HI and LO to E1364A NC line and mainframe chassis" 500 510 520 530 540 550 560 570 580 590 600 610 32 Verification Tests PAUSE DISP ! OUTPUT @Switch;"OPEN (@100:115)" !Close all NC contacts WAIT 1 OUTPUT @Dmm;"TRIG SGL" ENTER @Dmm;Rdg ! PRINT "DC Isolation -- Normally C
Performance Test Record Table 2-1, "Performance Test Record for the E1364A Switch Module," is a form you may copy and use to record performance verification test results for the switch module. Pages 3 and 4 of Table 2-1 show switch accuracy, measurement uncertainty (M.U.), and test accuracy ratio (TAR) values.
Table 2-1. Performance Test Record for the E1364A Switch Module (Page 1 of 4) Test Facility: Name______________________________________ Report No.
Table 2-1. Performance Test Record for the E1364A Switch Module (Page 2 of 4) Model ______________________________ Report No. ____________________ Date _________________ Test Equipment Used: Description Model No. Trace No. Cal Due Date 1. _______________________________ ______________ ______________ ______________ 2. _______________________________ ______________ ______________ ______________ 3. _______________________________ ______________ ______________ ______________ 4.
Table 2-1. Performance Test Record for the E1364A Switch Module (Page 3 of 4) Model ______________________________ Test No. Test Description Report No. _____________________________ Date_____________ Minimum* Measured Reading Maximum Meas Uncert TAR ____________ ____________ ____________ ____________ ____________ ____________ ____________ ____________ ____________ ____________ ____________ ____________ ____________ ____________ ____________ ____________ 3.5 3.5 3.5 3.5 3.5 3.5 3.5 3.5 3.5 3.5 3.
Table 2-1. Performance Test Record for the E1364A Switch Module (Page 4 of 4) Model ____________________________ Test No. Test Description Report No. _____________________ Date______________ Minimum Measured Reading Maximum* Meas Uncert TAR 6.0E6 6.0E6 6.0E6 6.0E6 NA NA NA NA 2-3. DC Isolation Test (Values in ohms) NO to Common NC to Common NO to chassis NC to chassis 1E8 1E8 1E8 1E8 ____________ ____________ ____________ ____________ *Single-sided specification -- Maximum does not apply.
Notes 38 Verification Tests Chapter 2
Chapter 3 Replaceable Parts Introduction This chapter contains information for ordering replaceable parts for the E1364A Switch Modules with serial number 2934A10370 and higher. The tables provide the following information: • Table 3-1 lists assembly and terminal module part numbers for the E1364A Switch Module. • Table 3-2 lists the Reference Designators for the switch module. Ordering Information To order a part listed in Table 3-1, specify the Agilent part number and the quantity required.
Table 3-2. Reference Designators A assembly PCB printed circuit board BRK bracket PNL panel C capacitor R resistor CR diode RP resistor pack LBL label RVT rivet F fuse SCR screw J electrical connector (jack) SHD shield JM jumper SW switch K relay TB terminal block (module) MP mechanical part U integrated circuit P electrical connector (plug) Figure 3-1.
Figure 3-2.
Notes 42 Replaceable Parts Chapter 3
Chapter 4 Service Introduction This chapter contains service information for the Agilent E1364A Switch Module, including troubleshooting techniques and repair and maintenance guidelines. WARNING Equipment Required Service Aids Chapter 4 Do not perform any of the service procedures shown unless you are a qualified, service-trained technician, and have read the WARNINGS and CAUTIONS in Chapter 1.
Troubleshooting Techniques To troubleshoot an E1364A Switch Module problem, you should first identify the problem and then isolate the cause to a replaceable part. See Chapter 3, "Replaceable Parts," or Appendix B, "Backdating Information," for descriptions and locations of replaceable parts, depending on the serial number of your switch module. Identifying the Problem Table 4-1 lists some common problems, along with symptoms and possible solutions.
Testing the Assembly You can use the tests and checks in Table 4-2 to isolate the problem. See Figure 3-1 in Chapter 3 or Figure B-1 in Appendix B for locations of replaceable parts, depending on the serial number of your switch module. Table 4-2. E1364A Tests/Checks Test/Check Reference Designator Check: Heat Damage ---------- Discolored PC boards Damaged insulation Evidence of arcing Switch/Jumper Settings JM13, JM14, ...
Repair and Maintenance Guidelines This section provides guidelines for repairing and maintaining the E1364A Switch Module, including: • ESD precautions • Soldering printed circuit boards • Post-repair safety checks ESD Precautions Electrostatic discharge (ESD) may damage static sensitive devices in the E1364A Switch Module. This damage can range from slight parameter degradation to catastrophic failure.
Appendix A Verification Tests - C Programs Functional Verification Test This program is designed to do the Functional Verification Test found in Chapter 2, "Verification Tests." Example This example sends a *IDN? command to the switch. This test can be used to verify that the switch is connected properly and is responding to a basic command. #include #include
Performance Verification Tests These programs are designed to do the Performance Verification Tests found in Chapter 2, "Verification Tests." Example: Closed Channel Resistance Test This example performs the Closed-Channel Resistance Test for all Normally Open and Normally Closed contacts. /* Closed-channel Resistance Test E1364A */ #include #include
iprintf (id, "CLOS (@%u)\n", channel); ipromptf (dm, "TRIG SGL\n", "%t", reading); printf ("\n Channel %u (closed): %s", i, reading); iprintf (id, "OPEN (@%u)\n", channel); ipromptf (dm, "TRIG SGL\n", "%t", reading); printf ("\n Channel %u (open): %s", i, reading); } printf ("\n\nConnect DMM to NC and C lines of E1364A (4-wire connection)"); getchar (); iprintf (id, "CLOS (@100:115)\n"); printf ("\n\nContact Resistance -- Normally closed contacts\n"); for (i = 0; i <=15; i++) { channel = 100 + i; iprintf (i
Example: DC Isolation Test This example performs the DC Isolation Test for Normally Open and Normally Closed channels to both Common and chassis (earth ground). /* DC Isolation Test E1364A */ #include #include
printf ("\n\nConnect DMM HI and LO to E1364A NO line and mainframe chassis"); getchar (); ipromptf (dm, "TRIG SGL\n", "%t", reading); printf ("\nDC Isolation -- Normally Open Channels to Chassis"); printf ("\n R = %s", reading); printf ("\n\nConnect DMM HI and LO to E1364A NC line and mainframe chassis"); getchar (); iprintf (id, "OPEN (@100:115)\n"); ipromptf (id, "*OPC?\n", "%t", reading); ipromptf (dm, "TRIG SGL\n", "%t", reading); printf ("\nDC Isolation -- Normally Closed Channels to Chassis"); printf
Notes 52 Example C Programs Appendix A
Appendix B Backdating Information Introduction This chapter contains information for ordering replaceable parts for the Agilent E1364A Switch Modules with serial numbers prior to 2934A10370. • Table B-1 lists the assembly and terminal module part numbers for the E1364A Switch Module. • Table B-2 lists the Reference Designators for the switch module. Ordering Information To order a part listed in Table B-1, specify the Agilent part number and the quantity required.
Reference Designator Part Number Qty Part Description ADDITIONAL COMPONENTS 3050-0082 2 WASHER-FL NM NO. 4 .116-IN-ID .188-IN-OD C1 0180-3899 1 CAPACITOR-FXD 220uF +-20% 50 V AL-ELCTLT C2 0160-3334 2 CAPACITOR-FXD 0.
Reference Designator Part Number Qty Part Description U25-U26 1820-5424 2 IC DRIVER CMOS/74HCT LINE OCTL U30 1820-4152 2 IC FF CMOS/74HCT D-TYPE POS-EDGE-TRIG U35 1820-3714 IC TRANSCEIVER TTL/ALS BUS OCTL U38 1820-4152 IC FF CMOS/74HCT D-TYPE POS-EDGE-TRIG E1364-80001 1 E1364A TERMINAL MODULE E1364-66510 1 TERMINAL CARD 16-CH GP RLY SWITCH E1300-84401 1 TERMINAL CARD, CASE ASSEMBLY Table 3-2.
Figure B-1.
Figure B-2.
Copyright © Agilent Technologies, Inc.