User's Manual

62 Chapter 2
Performance Verification Tests
6. Frequency Span Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B
Table 2-9 Frequency Span Readout Accuracy
Analyzer
Start
Frequency
(MHz)
Analyzer
Stop
Frequency
(MHz)
Synthesized
Signal
Generator
Frequency
(MHz)
Synthesized
Sweeper
Frequency
(MHz)
Test
Record
Entry
Analyzer
Span
(MHz)
0 3000 300 2700 1) 3000
10 110 20 100 2) 100
10 10.1 10.01 10.09 3) 0.1
800 900 810 890 4) 100
800 800.1 800.01 800.09 5) 0.1
1400 1500 1410 1490 6) 100
1499 1499.1 1499.01 1499.09 7) 0.1