   Using the Relay Multiplexer  39Chapter 2
Example: Switching 
Channels (Two-Wire)
This example illustrates two-wire mode operation. The HI and LO terminals 
of bank 0 channels 0 and 7 are closed, connecting them to the bank 0 HI and 
LO COM terminals. Figure 2-2 shows how the multiplexer is configured. 
For two-wire operation, the control relays are set as follows. 0990/0991 are 
opened if using the <channel_list> command with SCAN:PORT ABUS and 
SCAN:MODE <mode>. Mode can be RES, VOLT, or NONE. 0990/0991 are 
left in their present state if mode is FRES. 0992/0993 will close when 
SCAN:PORT ABUS is selected during a scan (see “Scanning Channels”). 
0994/0995/0996 remain in their present state with the following exceptions. 
0994 is closed in RES mode. If <card_number>, WIRE2X64 (2x64 
configuration), 0994 is closed in the RES and NONE modes. In the FRES 
mode, 0994 and 0995 are opened. 0996 closes and connects COM to LO 
for voltage measurements with the MEASure or SCPI commands in a 
scanning multimeter.
Figure 2-1. Example: Switching Channels (One-Wire)
Open
Closed
CH0991
CH0990
Closed
Open
CH0995
JM5JM4
JM2 JM3
H
L
COM
L
H
CH1
1W H
Cable T
1W L
1W Ref L
1 Wire High Common
Cable Test
1 Wire Low Common
1 Wire Low Ref
Bank 2
SWITCH CARD
TERMINAL MODULE