User manual
46 Using the Relay Multiplexer  Chapter 2
120 OUTPUT 70914;"TRIG:SOUR TTLT1"
!Multiplexer to advance scan when trigger 
received in TTL trigger bus line 1
130 OUTPUT 70914;"SCAN:MODE VOLT"
!Sets switchbox measurement to volt (used 
to make 2-wire resistance measurement on 
multimeter’s HI/LO terminals)
140 OUTPUT 70914;"SCAN:PORT ABUS"
!Closes control relays 992 and 993 connecting 
the analog bus to the upper and lower four 
bank commons
150 OUTPUT 70914;"SCAN (@100:177)"
!Defines channel list to scan bank 0 channel 
0 to bank 7 channel 7
160 OUTPUT 70914;"*OPC?"
!Multiplexer operations complete
170 ENTER 70914;Opc
!Enters a "1" when complete
180 OUTPUT 70903;"READ?"
!Places multimeter in wait-for-trigger state. 
Will send measurement results to output 
buffer when triggered.
190 OUTPUT 70914;"INIT"
!Closes bank 0 channel 0 and enables the 
scan. Causes a trigger output on TTL Trigger 
bus line 0 that initiates the multimeter to 
make a measurement.
200 ENTER 70903;Readings(*)
!Enters measurement results
210 PRINT Readings(*)
!Displays measurement result
220 END
!Terminates program
Example: Scanning 
Channels Using E1412A 
Multimeter 
This example program uses the E1406A Command Module for one-wire 
scanning of the E1460A using the two-wire ohms function of the E1412A 
multimeter. The TTL trigger bus lines perform E1412A multimeter triggering 
and E1460A multiplexer channel advance. This program uses the 
stand-alone switchbox mode. 
For this example, the following resistors are connected to the channels 
indicated and to the one-wire COM terminal. The remaining channels are 
open.
1 k on channel 00 HI
1.5 k on channel 00 LO
1.2 k on channel 01 HI
1.8 k on channel 01 LO
A typical result returns +0, "No error" for the response to the SYST:ERR? 
command. The four channels with resistors connected return the nominal 
value of the resistor, such as 1003.129 for channel 00 HI or 1489.102 for 
channel 00 LO. All other channels return 9.9E+37 to indicate an open 
channel.