56 Using the Relay Multiplexer  Chapter 2
Saving and 
Recalling States
The *SAV <numeric_state> command saves the current instrument state. 
The state number (0-9) is specified in the numeric_state parameter. The 
following settings are saved:
• Channel Relay States (bank 0-7 relays open or closed)
• Control Relay States (bank 9 relays open or closed)
• ARM:COUNt
• TRIGger:SOURce
• OUTPut[:STATe]
• INITiate:CONTinuous
• [ROUTe:]SCAN:MODE
• [ROUTe:]SCAN:PORT
The *RCL <numeric_state> command recalls a previously saved state. 
Enter the number (0-9) in the numeric_state parameter of the desired saved 
state. If *SAV was not previously executed using the selected number, the 
multiplexer module will configure to the reset values (refer to table 2-2).
NOTE *RCL, *RST, and *SAV do not affect the operating mode as set by the 
FUNCtion command or status register switch positions. If using the 
FUNCtion command to set the operating mode, the *SAV/*RCL command 
must be executed AFTER the FUNCtion command.
Detecting Error 
Conditions
There are two general approaches to error checking: polling and using 
interrupts. An example of each method follows. 
Example: Error Checking 
Using Polling
The simplest, but most time consuming, method is to ask the instrument 
whether there are errors at every step of the switching process. This is called 
polling and is illustrated in the following example.
10 DIM Err_num$[256]
20 OUTPUT 70914;"CLOS (@101)"
30 OUTPUT 70914;"SYST:ERR?"
40 ENTER 70914;Err_num$
50 IF VAL (Err_num$) 0 THEN
60 PRINT "Error";Err_num$
70 STOP
80 END IF
90 ...(program continues)