User manual
  Relay Multiplexer Command Reference  71Chapter 3
Parameters
 Comments Enabling TTL Trigger Bus: When enabled, a negative-going pulse is 
output from the selected TTL Trigger bus line (0 to 7) after each channel 
in the switchbox is closed during a scan. If disabled, a pulse is not output. 
TTL Trigger Bus Line Shared by Switchboxes: Only one switchbox 
configuration can use the selected TTL Trigger at a time. When enabled, 
the selected TTL Trigger bus line (0 to 7) is pulsed by the switchbox each 
time a scanned channel is closed. To disable the output for a specific 
switchbox, send the OUTPut:TTLTrgn OFF or 0 command for that 
switchbox.
One Output Selected at a Time: Only one output (ECLTrg0 or 1; TTLTrg0, 
1, 2, 3, 4, 5, 6>, or 7; or EXTernal) can be enabled at one time. Enabling 
a different output source will automatically disable the active output. 
For example, if TTLTrg1 is the active output and TTLTrg4 is enabled, 
TTLTrg1 will become disabled and TTLTrg4 will be the active output.
Related Commands: [ROUTe:]SCAN, TRIGger:SOURce, 
OUTPut:TTLTrgn[:STATe]?
*RST Condition: OUTPut:TTLTrgn[:STATe] OFF (disabled)
Example Enabling TTL Trigger Bus Line 7
OUTP:TTLT7:STAT 1
!Enable TTL Trigger bus line 7 to output 
pulse after each scanned channel is 
closed
OUTPut:TTLTrgn[:STATe]?
OUTPut:TTLTrgn[:STATe]? queries the present state of the specified TTL 
Trigger bus line. The command returns "1" if the specified TTLTrg bus 
line is enabled or "0" if disabled.
Example Query TTL Trigger Bus Enable State
This example enables TTL Trigger bus line 7 and queries the enable 
state. The OUTPut:TTLTrgn? command returns "1" since the port is 
enabled.
OUTP:TTLT7:STAT 1
!Enable TTL Trigger bus line 7
OUTP:TTLT7?
!Query bus enable state
Name Type Range of Values Default
 <n> numeric 0 or 7  N/A
 <mode> boolean 0 I 1 I OFF I ON  OFF I 0