  Relay Multiplexer Command Reference  75Chapter 3
[ROUTe:]FUNCtion
[ROUTe:]FUNCtion <card_number>, <function> selects the operating 
mode of the multiplexer channels. All channels on the card specified by 
card_number operate in the specified mode. [ROUTe:] is NOT optional 
when ROUT:FUNC is used with a scanning multimeter configuration.
Parameters
Comments  ROUTe is Not Always Optional: If used with a scanning multimeter 
configuration, ROUTe:FUNCtion must be used.
Command Not Always Used: This command is not required if the status 
register switch is configured to the desired mode.
Using the FUNCtion Command: When using the FUNCtion command to 
reconfigure the multiplexer to a different operating mode than the status 
register switch is set to, the command must be sent 
AFTER the card is 
powered up. 
One-Wire Mode (WIRE1): Switches either the HI or LO terminal of a 
channel in banks 0 through 7, to the one-wire HI COM or one-wire LO 
COM terminal. When closing a channel in one-wire mode, the HI or LO 
line must be selected using channel_list. Only one of the 128 one-wire 
channels can be switched at a time.
Two-Wire Mode (WIRE2): Switches both the HI and LO terminals of a 
channel in banks 0 through 3 or banks 4 through 7 to that bank’s HI COM 
and LO COM terminals.
Two-Wire Mode (WIRE2X64): Switches the HI and LO terminals of a 
channel in banks 0 through 7 to that bank’s HI COM and LO COM 
terminals. A maximum of 64 two-wire channels can be switched. This 
mode is available via E1406A (Switchbox Rev. A06.00 or later). Prior to 
this revision, closing control relay 0995 in two-wire mode will change the 
card configuration to a single 64-channel two-wire multiplexer.
Three-Wire Mode (WIRE3): Banks are paired 0/4, 1/5, 2/6, and 3/7. 
Switches both the HI and LO terminal of a channel in bank 0-3, to that 
bank’s HI and LO COM terminals. Also switches the LO terminal of the 
channel in pair bank 4-7 to that bank’s LO COM terminal. 
Name Type Range of Values Default
<card_number> numeric 01 to 99 N/A
<function> discrete WIRE 1 I WIRE 2 |
WIRE 2X64 I 
WIRE 3 I WIRE 4
WIRE 2