AUTO-SERIES OPERATION
Auto-series operation permits equal or proportional volta
g
e
sharin
g
, and allows control of output volta
g
e from one master
unit. The volta
g
e of the slaves is determined b
y
the settin
g
of
the front panel VOLTAGE control on the master and volta
g
e
divider resistor. The master unit must be the most positive sup-
pl
y
of the series. The output CURRENT controls of all series
units are operative and the current limit is equal to the lowest
settin
g
. If an
y
output CURRENT controls are set too low, auto-
matic cross over to constant current operation will occur and the
output volta
g
e will drop. Fi
g
ure 12 and Fi
g
ure 13 show the rear
panel switch settin
g
s and terminal connections for Auto-series
operation of two supplies and three supplies. This mode can
also
g
ive ±volta
g
e trackin
g
operation of two supplies with two
separate loads.
Mixed model numbers ma
y
be emplo
y
ed in auto-series combi-
nation without restriction, provided that each slave is specified as
bein
g
capable of auto-series operation. If the master suppl
y
is set
up for constant current operation, then the master-slave combina-
tion will act as a composite constant current source.
Total output volta
g
e to
g
round must not exceed 240 Vdc.
Determinin
g
Resistors.
External resistors control the fraction (or
multiple) of the master unit's volta
g
e settin
g
that is supplied from
the slave unit. Notice that the percenta
g
e of the total output volt-
a
g
e contributed b
y
each suppl
y
is independent of the ma
g
nitude
of the total volta
g
e. For two units in auto-series the ratio of R1 to
R2 is
(R1+R2)/R1 = (Vo/Vm)
R2/R1 = (Vs/Vm)
Where Vo = auto-series volta
g
e = Vs + Vm
Vm = master unit's output volta
g
e
Vs = slave unit's output volta
g
e
For example, usin
g
the E3617A as a slave unit and puttin
g
R2=50
k (1/4 watt), then from the above equations,
R1 = R2(Vm/Vs) = 50(Vm/Vs) k
In order to maintain the temperature coefficient and stabilit
y
perfor-
mance of the suppl
y
, choose stable, low noise resistors.
It is recommended to connect a 0.1 µF capacitor in paral-
lel with R2 in two supplies operation or R2 and R4 in
three supplies operation to ensure the stable operation.
Settin
g
Volta
g
e and Current.
Use the master unit's controls to
set the desired output volta
g
e and current. The VOLTAGE control
of the slave unit is disabled. Turnin
g
the volta
g
e control of the
master unit will result in a continuous variation of the output of the
series combination, with the contribution of the master's output
volta
g
e to that of the slave's volta
g
e alwa
y
s remainin
g
in the ratio
of the external resistors. Set the CURRENT control of slave unit
above the master unit's current settin
to avoid havin
the slave
switch to CC operation.
When in CC operation the combined output current is the same
as the master unit's current settin
, and when in CV operation the
combined output volta
e is the sum of the master unit's and the
slave unit's output volta
es.
Overvolta
e Protection.
Set the OVP shutdown volta
g
e in each
unit so that it shuts down at a volta
e hi
her than its output volta
e
durin
auto-series operation. When a master unit shuts down, it pro-
rams an
slave units to zero output. When a slave unit shuts down,
it shuts down onl
itself (and an
slaves below it in the stack). The
master (and all slaves above the shut-down slave) continues to sup-
pl
output volta
e.
Fi
g
ure 12. Auto-Series Operation of Two Supplies
Fi
g
ure 13. Auto-Series Operation of Three Supplies
MASTER
SLAVE
CV CC SENSE
LOCAL
REMOTE
OUT
+S -S
+
_
CV CC
VREF
A1 A2 A3 A4 A5
+
+
M/S 1 M/S 2
__
MASTER
SLAVE
CV CC SENSE
LOCAL
REMOTE
OUT
+S
-S
+
_
CV CC
VREF
A1 A2 A3 A4 A5
+
+
M/S 1 M/S 2
_
_
MASTER POWER SUPPLY
LOAD
SLAVE POWER SUPPLY
R1 R2
MASTER
SLAVE
CV CC SENSE
LOCAL
REMOTE
OUT
+S
-S
+
_
CV CC
VREF
A1 A2 A3 A4 A5
+
+
M/S 1 M/S 2
__
MASTER POWER SUPPLY
LOAD
MASTER
SLAVE
CV CC SENSE
LOCAL
REMOTE
OUT
+S
-S
+
_
CV CC
VREF
A1 A2 A3 A4 A5
+
+
M/S 1 M/S 2
__
SLAVE POWER SUPPLY(S1)
MASTER
SLAVE
CV CC SENSE
LOCAL
REMOTE
OUT
+S
-S
+
_
CV CC
VREF
A1 A2 A3 A4 A5
+
+
M/S 1 M/S 2
__
SLAVE POWER SUPPLY(S2)
R1 R2
R3 R4
Vo=Vm(1+
R2
R1
R2
R1
+
R4
R3
)
Where Vo = Auto-Series volta
g
e = Vm + Vs1 + Vs2
Vm = master unit's output volta
g
e
Vs1 = slave(S1) unit's output volta
g
e
Vs2 = slave
(
S2
)
unit's output volta
g
e