Technical data

1 Calibration Procedures
Recommended Test Equipment
4 E3632A Service Guide
Recommended Test Equipment
The test equipment recommended for the performance
verification and adjustment procedures is listed below. If the
exact instrument is not available, use the accuracy
requirements shown to select substitute calibration
standards.
Table 1 - 1 Recommended test equipment
Instrument Requirements Recommended model Test function
GPIB controller Full GPIB capabilities Agilent 82341C Interface card Programming and readback
accuracy.
Oscilloscope 100 MHz with 20 MHz
bandwidth
Agilent 54602B Display transient response
and ripple and noise
waveform.
RMS voltmeter 20 MHz Measure rms ripple and
noise.
Digital voltmeter Resolution: 0.1 mV
Accuracy: 0.01%
Agilent 34401A Measure DC voltages.
Electronic load Voltage Range: 50 Vdc
Current Range: 10 Adc
Open and Short Switches
Transien t On /O ff
Agilent 6063B Measure load and line
regulations and transient
response time.
Resistive loads (R
L
)2.1 Ω, 200 W
7.5 Ω, 200 W
Measure ripple and noise.
Current monitoring resistor
(shunt)
0.01 Ω, 0.01% Constant current test setup.