Agilent E5052BU Option 018 Hard Disk Drive Kit User’s Guide Third Edition Manufacturing No.
Notices The information contained in this document is subject to change without notice. This document contains proprietary information that is protected by copyright.All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of the Agilent Technologies. © Copyright 2007, 2008, 2009 Agilent Technologies Manual Printing History The manual’s printing date and part number indicate its current edition.
1 General Information This chapter describes the Hard Disk Drive Kit, contents of the kit, and required tools to install the Option 018 into the Agilent E5052B.
General Information Option 018 Hard Disk Drive Kit Option 018 Hard Disk Drive Kit The purpose of this kit is to install the option 018 into the Agilent E5052B. Contents of the Agilent E5052BU Option 018 Hard Disk Drive Kit The kit includes the parts listed in Table 1-1. If anything is missing, contact the nearest Agilent Technologies office. Table 1-1 Contents of the Agilent E5052BU Option 018 Hard Disk Drive Kit Description Qty.
2 Option Installation This chapter describes the procedure to install the option 018 into the Agilent E5052B.
Option Installation Introduction Introduction This chapter describes the procedure to install Option 018 into the E5052B. The procedure consists of the following items.
Option Installation Procedure Saving System calibration Data from Removable Hard Disk to USB Pen Drive The following is a description of the procedures to save the system calibration data stored in the removable hard disk to a USB Pen drive. Step 1. Exit the measurement window. System > Service Menu > Administorator Menu > Exit E5052B > OK Step 2. Insert a USB Pen drive into the USB port in front panel. Step 3. Open D:\Recovery drive. Step 4. Click .D:\syscal from the explorer. Step 5.
Option Installation Procedure Removing Removable Hard Disk Follow the steps below to remove a removable hard disk. CAUTION Before removing a hard disk, you must turn off the unit’s power and then disconnect the power cord from the outlet. Step 1. Turn off the power. Step 2. Insert the key in the lock and turn anti-clockwise. Step 3. Open the screw in clockwise direction. Step 4. Press the button under direction. Step 5. Remove the hard disk.
Option Installation Procedure Recalling System Calibration Data from USB Pen drive to Removable Hard Disk The following is a description of the procedures to copy the system calibration data stored in the USB Pen drive to the removable hard disk. Step 1. Exit the measurement window. System > Service Menu > Administorator Menu > Exit E5052B > OK Step 2. Insert a USB pen drive in to the USB port in front panel. Step 3. Open D:\Recovery drive. Step 4.
Option Installation System Recovery System Recovery By executing system recovery, you can return the system of the E5052B (the Windows operating system and the firmware) to the factory state (at the time of purchase*1). NOTE System recovery is a function of Windows and not a function of the E5052B Firmware. The system recovery procedure is different in accordance with the HDD revision. The HDD revision can be confirmed using the following procedure: Step 1.
Option Installation System Recovery Notes on executing the factory recovery function Executing the factory recovery function causes the following conditions: R In addition to the Windows operating system and the firmware, the following settings of the E5052B are returned to the factory state. • • • Network setting GPIB setting Printer setting R The driver for the supported printer installed after purchase is deleted. R You need to execute initial registration again.
Option Installation System Recovery Procedure to execute the factory recovery (HDD Revision: NM321 or above) NOTE The recovery function for Factory image is only loaded. The User back up function is not available. Step 1. Turn ON the E5052B. Step 2. The boot option is displayed. Use the downward arrow key to select the “Agilent Recovery Solution” and press Enter Step 3. Windows PE starts up and “Agilent Recovery Program” window is displayed. Press 1, and then press Enter.
Option Installation System Recovery Figure 2-3 Agilent Recovery Program(1/3) Step 6. Recovery function is performed. Step 7. A message box appears after the completion of recovery process. Press any key to continue. Figure 2-4 Agilent Recovery Program(1/4) Step 8. When factory recovery is finished, E5052B restarts. Step 9. After restart, the screen for initial registration appears. Execute initial registration.
Option Installation System Recovery Procedure to execute the factory recovery (HDD Revision: NM320 or below) This section describes how to return the contents of the C drive to the factory state. NOTE You need a keyboard for this operation. Step 1. Shut down the E5052B. Step 2. Disconnect all of the USB device from the USB ports. Step 3. Connect the keyboard to the E5052B. Step 4. Press the standby switch of the E5052B to turn it on. Step 5.
Option Installation System Recovery Step 6. Select Restore. Figure 2-5 Activate Acronis Startup Recovery Manager e5071cse1079 Step 7. A confirmation dialog box is displayed. Push the enter to select Yes to start the recovery of the factory backup. The recovery takes a few minutes depending on the amount of data. Figure 2-6 Recover Factory Backup e5052bse1018 Step 8. When the recovery is finished, push the enter select OK. The E5052B restarts automatically. Step 9.
Option Installation System Recovery CAUTION Never turn off the power during the system recovery because doing so may cause serious damage to the E5052B. Procedure to create the user backup image (only available when HDD Revision is NM320 or below) This section describes how to create the user backup image. The C and F drives’ contents saved in this procedure are recalled when the user recovery function is executed. Step 1.
Option Installation System Recovery Step 5. Click Backup e5071cse1125 Step 6. Click Next button.
Option Installation System Recovery Step 7. Select “The entire disk contents or individual partition,” then click Next button. e5071cse1127 Step 8. Select the C drive and F drive. e5052bse1019 Step 9. Select the destination folder and input a file name, then click Next button. It is recommended to save the file to a directory which has enough space like the D drive.
Option Installation System Recovery never save the contents of the drive to the C, E or F drive. e5052bse1020 Step 10. Select “Set the options manually” and click Next button. e5071cse1130 Step 11.
Option Installation System Recovery button. e5071cse1131 Step 12. If necessary, input comments and click Next button. The comments can be omitted.
Option Installation System Recovery Step 13. Click Proceed button. e5052bse1021 Step 14. The following screen appears during backup. e5052bse1022 Step 15. The following screen appears when the user backup image was created. Click OK.
Option Installation System Recovery CAUTION Never turn off the power during creating the user backup image because doing so may cause serious damage to the E5052B.
Option Installation System Recovery Procedure to execute the user recovery function (only available when HDD Revision is NM320 or below) Returns the contents of the C and F drives to a user-specified state. To use this function, you must create the user backup image in advance. For more information, see the description “Procedure to create the user backup image (only available when HDD Revision is NM320 or below)” on page 16. Step 1. Close the E5052B application by pressing System > Service Menu > Exit.
Option Installation System Recovery Step 6. Click Next button. e5071cse1137 Step 7. Select the saved backup image file, then click Next button.
Option Installation System Recovery Step 8. Select “Restore disks or partitions,” then click Next button. e5071cse1139 Step 9. Select the C drive, then click Next button.
Option Installation System Recovery Step 10. Select the C drive to restore, then click Next button. e5052bse1026 Step 11. Select “Active” and click Next button.
Option Installation System Recovery Step 12. Click Next button. e5052bse1027 Step 13. Select “Yes, I want to assign a logical drive letter to the restored partition.” and click Next button.
Option Installation System Recovery Step 14. Select “Yes, I want to restore another partition or hard disk drive.” then click Next button. e5052bse1077 Step 15. Select the F drive to restore, then click Next button.
Option Installation System Recovery Step 16. Select the F drive to restore, then click Next button. e5052bse1029 Step 17. Select “Logical” and click Next button.
Option Installation System Recovery Step 18. Click Next button. e5052bse1030 Step 19. Select “Yes, I want to assign a logical drive letter to the restored partition,” then click Next button.
Option Installation System Recovery Step 20. Select “No, I do not.” and click Next button. e5052bse1079 Step 21. Select “Set the options manually” and click Next button.
Option Installation System Recovery Step 22. Select “Restoration Priority” in the left field and “High” in the right field, then click Next button. e5052bse1081 Step 23. Click Proceed button. e5052bse1032 Step 24. Click Reboot button.
Option Installation System Recovery CAUTION Never turn off the power during the recovery of the user backup image because doing so may cause serious damage to the E5052B. Step 25. After boot up, perform “Initial Registration of E5052B” on page 34 and “Calibration of the Touch Screen” on page 39.
Option Installation Initial Registration of E5052B Initial Registration of E5052B When you start up the E5052B at the first time, you need to perform the initial registration of the Windows XP operating system of the E5052B. There are two procedures in accordance with the HDD revision. NOTE It is not possible to check the HDD revision installed on your E5052B before Windows is booted. Therefore, you should follow the procedure which you see on the unit.
Option Installation Initial Registration of E5052B Step 3. In the Windows XP Professional Setup dialog box, read the license agreement, then select the I accept this agreement box and click the Next > button (Figure 2-9). Figure 2-9 Windows XP Professional Setup dialog box (1/2) e5071cse1074 Step 4. In the next dialog box, input E5052B in the Name box. Then, click the Next> button (Figure 2-10). Figure 2-10 Windows XP Professional Setup dialog box (2/2) E5052 e5052bse1033 Step 5.
Option Installation Initial Registration of E5052B Figure 2-11 Agilent Recovery Image Setup Screen(1/1) Step 7. A progress bar appears (Figure 2-12) indicating the progress of the Recovery Image setup. The complete process takes less than 5 minutes to finish. Figure 2-12 Agilent Recovery Image Setup Screen(1/2) Step 8. A message dialog box appears after the completion of the Recovery Image Setup procedure. The dialog box disappears automatically after a few seconds. Step 9.
Option Installation Initial Registration of E5052B Initial Registration Procedure (HDD Revision: NM320 or below) Step 1. Turns on the E5052B. Step 2. The Windows XP Professional Setup wizard appears. Click the Next > button (Figure 2-13). Figure 2-13 Windows XP Professional Setup wizard e5071cse1073 Step 3. In the Windows XP Professional Setup dialog box, select the I accept this agreement box and click the Next > button (Figure 2-14).
Option Installation Initial Registration of E5052B Figure 2-15 Windows XP Professional Setup dialog box (2/2) E5052 e5052bse1033 Step 5. After a while, the E5052B restarts automatically. Then, the measurement display appears.
Option Installation Calibration of the Touch Screen Calibration of the Touch Screen After E5052B measurement screen appears, you have to calibrate the touch screen. Follow the procedure described below to calibrate the touch screen. Step 1. Press . Step 2. Press Service Menu. Step 3. Press Test Menu. Step 4. Press Adjust Touch Screen. The touch screen calibration screen (Figure 2-16) appears. Figure 2-16 Touch Panel Calibration Screen e5071cse1078 Step 5.
REGIONAL SALES AND SUPPORT OFFICES For more information about Agilent Technologies test and measurement products, applications, services, and for a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir. You can also contact one of the following centers and ask for a test and measurement sales representative.