Installation guide
Chapter 2  15
Installation
Environmental Requirements
2. Installation
Ventilation Requirements
To ensure safety requirements, the specifications and measurement accuracy of the 
analyzer, you must keep environmental temperature within the specified range by 
providing appropriate cooling clearance around the analyzer or, for the rackmount type, by 
forcefully air-cooling inside the rack housing. For more information on environmental 
temperature to satisfy the specifications and measurement accuracy of the analyzer, see the 
specification in the E5071C Online Help.
When the environmental temperature around the analyzer is kept within the temperature 
range of the operating environment specification (See Section 
“Operating Environments” 
on page 14), the analyzer conforms to the requirements of the safety standard. 
Furthermore, under that temperature requirement, the analyzer still conforms to the 
requirements of the safety standard even when placing the analyzer with cooling clearance 
as follows:
Requirements
Back ≥ 180 mm
Sides ≥ 60 mm (both right and left)
Figure 2-1  Ventilation space at the installation
e5071cqe1001