Installation guide

1-38
Agilent Technologies E5500B Installation Guide
Welcome to the E5500B Phase Noise Measurement System Series of
Solutions
3. From the Test Set pull-down list, select Agilent/HP 70420A. Click the
Check I/O button. A green check-mark will appear after the I/O check
has been performed by the software. If a red circle with a slash appears,
return to the Asset Manager (click the Asset Manager button) and
verify that the Agilent/HP 70420A and PC-Digitizer are configured
correctly (check that the license key has been entered correctly). Also
check your system hardware connections. Click the Check I/O button
for a re-check.
4. From the FFT Analyzer pull-down list, select II PCI20428W-1.Click
the Check I/O button. A green check-mark will appear after the I/O
check has been performed by the software.
5. From the Swept Analyzer pull down list select E4411A. Click the
Check I/O button. A green check mark will appear after the I/O is
completed.
Selecting the Agilent/HP 70420A test set, the E4411A and the
II PCI20428 PC-digitizer will tie those assets to the confidence test we
will be performing in the next step.
Click the Close button.