User guide

38
Fragment Library Preparation Using the AB Library Builder
System: 5500 Series SOLiD
Systems User Guide
Chapter5 Troubleshooting
Instrument error codes
Instrument error codes
If an extraction run is interrupted by an error, you cannot resume the interrupted run.
Follow the procedure below to resolve the error before you start a new run.
If you observe an error code:
1. Make a note of the error code, including the line number. Common error codes
are listed in the following table:
2. Press ESC to return to the Main menu.
3. If there are tips attached to the nozzles, press 1 to select the Manual screen, then
press 2 to return the tips to the original position.
4. Power OFF the instrument, remove the protocol card, wait 5 minutes, insert the
protocol card, then power on the instrument.
5. Run the axis test (refer to the AB Library Builder
System User Guide).
6. If the axis test:
Is successful, start a new extraction run. Use new samples and plastics where
required.
•Is not successful, contact Technical Support (“Obtaining support” on page
81).
Code Problem Code Problem
10 Failed return to origins, protocol cannot run 22 M axis time out, protocol in run
11 Limit error, protocol can not run 23 Y axis time out, protocol in run
12 Failed to return to Z Axis, protocol in run 24 Open door in motion
13 Failed to return to P axis, protocol in run 25 Abnormal input from bottom sensor in motion
14 Failed to return to M axis, protocol in run 26 Failed to initialize heating block
15 Failed to return to Y axis, protocol in run 27 Failed to initialize motion control board
16 Z axis limit error, protocol in run 110 System error; (Assigned greater than 10)
19 Y axis end limit, protocol in run
20 Z axis time out, protocol in run
21 P axis time out, protocol in run