User`s guide

Read This First
N9410-90001 Agilent 5500 SPM User’s Guide 4
Specifications
Environmental Conditions
Temperature (Operating): 5 to 40 °C
Temperature (Non-operating): -40 to 70 °C
Relative Humidity (Operating): 15 to 95 % non-condensing
Altitude: 2000 m
Power Requirements
100/120/220/240 VAC, 50/60 Hz
Mains supply voltage fluctuations are not to exceed 10 % of the nominal
supply voltage.
Equipment Class I, Pollution Degree 2, Installation Category II.
This equipment is for indoor use only.
When the product is subjected to 8 kV air discharge or 4 kV contact
discharge in accordance with IEC 61000-4-2, interruption of the laser
output may occur.
If this happens, laser power must be re-cycled in order to resume normal
operation. CAUT
Laser Safety Information
This system is designed to be used with a Class II or Class III diode
laser with an output of up to 1 mW of visible radiation at 670 nm or
980 nm. The aperture in the AFM scanning head is labeled with the
NOTE
These specifications apply to the Agilent 5500 system, and do not
guarantee the function of an experiment (including the cantilever) in
these conditions.
CAUTION
Stop using the scanner if the scanner cable insulation is damaged in order
to avoid electrical shock. Have it repaired or replaced by the factory.