User`s guide

Read This First
N9410-90001 Agilent 5500 SPM User’s Guide 5
logotype (shown below). DO NOT stare directly into the laser beam. To
ensure safe operation, the scanner must be operated and maintained in
accordance with the instructions included with the laser. The laser must
only be powered by a controller that includes an on/off switch, such as
the Agilent SPM Controller. DO NOT attempt to make any adjustments
to the laser, the lasers electronics, or optics. If laser malfunction is
suspected, immediately return the scanner to Agilent Technologies,
Inc.for repair or replacement; there are no user-serviceable parts. WA
RN
In accordance with federal FDA requirements, one of the following
laser precautions is affixed to the scanner:
Power Supply
It is not necessary to open the Agilent AFM Controller to make changes
to the power supply. However, the power cord should always be
unplugged before making any adjustments to the power source. The
Agilent SPM Controller has several different power supply options.
Procedure for Changing Input Voltage
1 Unplug the power cord from the Agilent AFM Controller.
2 Remove the fuse holder located on the back of the controller.
3 Underneath where the fuse holder was located is the input voltage
control switch. Pull out the switch and rotate it to the desired input
voltage 100/120/220/240 V.
4 Reinsert the voltage switch with the desired voltage.
5 Replace the fuse holder.
WARNING
Use of controls or adjustments or performance of procedures other than
those specified herein may result in hazardous light exposure.
Furthermore, the use of optical instruments with this product may
increase eye hazard.