User`s guide

Setting Up the Agilent 5500 SPM 3
Agilent 5500 SPM User’s Guide 58
Connecting the Components
The cabling for the standard 5500 SPM is shown in Figure 31. Other
cabling configurations are included in Appendix A.
Figure 31 Cabling for basic 5500 SPM configuration
Guidelines for Moving the System
Should you ever need to relocate the system, here are important
guidelines which must be followed to ensure safe operation:
The new location must meet all of the facility specifications
described above.
Turn off all components before disconnecting cables.
Disconnect all cables before moving any components.
Remove the scanner, detector and sample from the 5500 SPM before
moving the microscope base.
Remove the microscope base from the vibration isolation chamber
and transport both separately.
Follow the cabling diagrams exactly, being sure to connect all cables
before powering up the components.
CAUTION
Always make sure that all cables are connected before turning on any of
the components. Failure to do so can result in damage to equipment.