Register Definitions B
E6198B Switch/Load Unit User Manual B-23
Agilent U7177A 24-Channel Medium-Current Load Card With Current
Sense
The Agilent U7177A Load Card is designed for moderate
current loads mounted outside the load card. The card provides
current sense, pull-up/down selection, Card Type, Card
Configuration, and Status readback of the built in relay timers.
See “Using the Agilent U7177A 24-Channel Load Card" for more
information.
Register definitions for the card are as follows:
Card Type (R) Base + 0h
This register reads back the Card Type (18h) of the card.
Read Only
Power On/Reset State = 18h
Card Configuration (R) Base + 1h
This register reads back the Card Configuration. The Card
Configuration is determined by the user to distinguish different
load configurations of the same load card Card Type.
Read Only
Undefined bits readback as all 1s
Status (R) Base + 2h
This register reads back the card's status. Currently the status is
defined as the state of the relay timer. The relay timer has two
timers wire-OR'd together. One timer is designed for the slower
armature relays (>8 ms) and the second designed for the faster
reed relays (>500
μs). The timers restart whenever the registers
Bit 7-0
Purpose
Card Type
Setting 18h
Bit 7-0
Purpose
Card Configuration
Setting state