Technical data
22 Getting Started Guide
1 Quick-Start
Instrument Information
Power requirements
The only physical installation of your Agilent test set is a connection to a power source. 
Line voltage does not need to be selected. 
This test set does not contain customer serviceable fuses.
This is a Safety Class 1 Product (provided with a protective earthing 
ground incorporated in the power cord). The mains plug shall only be 
inserted in a socket outlet provided with a protective earth contact. 
Any interruption of the protective conductor inside or outside of the 
product is likely to make the product dangerous. Intentional 
interruption is prohibited. (IEC 348 clauses 17.3.3c & 17.3.4)
Failure to ground the test set properly can result in personal injury. 
Before turning on the test set, you must connect its protective earth 
terminals to the protective conductor of the main power cable. Insert 
the main power cable plug into a socket outlet that has a protective 
earth contact only. DO NOT defeat the earth-grounding protection by 
using an extension cable, power cable, or autotransformer without a 
protective ground conductor.
This product is designed for use in Installation Category II and Pollution 
Degree 2 per IEC 61010 Second Edition and IEC 664 respectively.  
This instrument has autoranging line voltage input. Before switching on 
the instrument, be sure the supply voltage is within the specified range.
The Mains wiring and connectors shall be compatible with the connector 
used in the premise electrical system. Failure, to ensure adequate earth 
grounding by not using the correct components may cause product 
damage, and serious injury.