Technical data
 14 Chapter 1 
What is the Sequence Analyzer?
Sequence Analyzer Introduction
What is the Sequence Analyzer?
Source List Sequencer
The signal generator (on the source side of the test set) is used to produce RF stimulus waveforms for use 
in calibrating and testing mobile devices. The Source List Sequencer can be used to execute a series of 
stimulus operations, in either of two ways:
• In the Sequence Analyzer mode, the source list sequencer is run by the same sequence which is also 
running the analyzer list sequencer. (This is the only way to run both list sequencers at the same 
time.) In this case, the lines representing the source sequence parameters are added to the sequence 
following the analyzer sequence parameters, as illustrated below in Figure 1-4.)
• Outside the Sequence Analyzer mode, the source list sequencer can be run independently of the 
analyzer list sequencer. (This is can be done from the front panel of the test set by selecting “List 
Sequencer” on the Source menu.) 
Figure 1-4.  Source Parameters Within a Combined Sequence
For example, two stimulus steps would be represented by two lines of parameters in the sequence file 
(tabs are represented here by “>”):
### Source ###
1 > IMM > 1 > PGSM > UP > 30 > -55 > gmsk.wfm > TIME > 4.62 > NONE 
2 > IMM > 1.33 > PGSM > UP > 31 > -35 > CONT > COUNt > 13 > NONE
The parameters for the source are described in Chapter 3 . For more detailed information on setting up 
the source List Sequencer refer to the Source User’s Guide, which is available as a pdf on the test set and 
on the Agilent web.