Example Test Scenarios
Verification Example
Example Test Scenarios
To add measurements to the acquisition, we click Edit Analysis Steps and add steps to the table. For the 
selected step (highlighted in the table), we select a measurements from the checkboxes (
Basic Tx Power 
in this case); the 
Meas Bitmap column in the table is automatically updated to reflect our selection of 
measurements. The 
Analysis Offset, Analysis Interval, and Expected Power values can be entered in 
the table directly, but it is more convenient to adjust these values graphically, using the slider bars on the 
display; the blue bars representing the analysis intervals on the screen change position and width 
accordingly.
Figure 11-22.  Adding Analysis Intervals