Mode Parameters
Test Set Parameters
Mode Parameters
Test Set Parameters
In addition to the parameters described early in this chapter, there are also parameters that are used to 
configure the test set for a sequence. These parameters can be set up from the front panel or by remote 
control SCPI commands. For details on how to set up these parameters from the front panel see Chapter 
6 , “Front Panel Sequence Setup,” on page 77.
Attenuation
In the test set you can set a mechanical and/or electronic attenuation. The electronic attenuation is set in 
the Acquisition setup, but the mechanical attenuation must be set prior to running the Sequence. The 
mechanical attenuation is set from the front panel by selecting the 
AMPTD Y Scale hardkey and 
selecting 
Attenuation. The remote control command for attenuation is:
[:SENSe]:LSEQuencer:POWer[:RF]:ATTenuation <rel_ampl>
This command is documented in more detail in the Sequence Analyzer online help (or in pdf format, the 
User’s and Programmer’s Reference).
RF inputs and outputs
There are four RF connectors on the left side of the test set’s front panel, which can be configured as 
follows:
•
RF Input can be selected as the input port, but not as the output port.
•
RF Output can be selected as the output port, but not as the input port.
•
RF I/O 1 can be selected as the input port, or the output port, or both.
•
RF I/O 2 can be selected as the input port, or the output port, or both.
NOTE It is not possible to have more than one input port at a time, or more than one output port 
at a time. However, it is possible to use one of the RF I/O connectors as the input port and 
the output port simultaneously.