Technical data
 92 Chapter 7 
List Sequencer Spreadsheet Data Entry
Tab separated value file
List Sequencer Spreadsheet Data Entry
Tab separated value file
A tab separated value (tsv) file can be used to store the settings associated with both analyzer and source 
sequences. There are two ways to generate the tsv file. You can set up the source and analyzer sequences 
using the front panel (see Chapter 6 ) and then save the sequence settings using the Save front panel 
hardkey. The values you have entered are saved into a tsv file. Alternatively, you can use a spreadsheet to 
generate a tsv file. The tsv file generated from the spreadsheet is then saved to the D-drive on the test set 
in My Documents\Sequences and can be used to set up the List Sequencers.
The tsv file generation can also be done remotely using SCPI commands.
This chapter shows you how to enter data into a spreadsheet to quickly and easily generate a tsv file for 
use in the list sequencers. A specially designed spreadsheet template is available on the Agilent web or 
simply use the guidelines below to make your own. You can take a copy of the spreadsheet template and 
enter values for the Sequence, Acquisitions, Analysis Intervals and source Sequences and Steps as 
required. 
The tsv file which you generate should be stored in My Documents\Sequences on D-drive of the test set, 
in order to use it in the list sequencers. To import the tsv file into the list sequencers, either specify the 
file name as a SCPI command parameter or from the front panel use the Recall hardkey menu.
If you set the list sequencers up from the front panel, you can save your sequence settings using the Save 
hardkey menu (see “Save and Recall” on page 75).










