Technical data
100 Chapter 3
1xEV-DO Analyzer and Over Air Test
Optional 1xEV-DO Tx and Over Air Analyzer Measurement Settings
1xEV-DO Analyzer and Over Air
Test
Setting the Maximum Inactive Channel Hight Limit
Step Notes
1 Press [Max Inactive Ch
Limit].
2 Press [Limits]. Enables you to toggle the Maximum Inactive
Channel measurement’s high limit function
On
(active) and
Off (inactive).
• If set to
On, the displayed Maximum Inactive
Channel metrics will change from blue to red
(fail), or to green (pass), and will indicate either
“P” (pass) or “F” (fail). If the measured value
falls above the high limit, the words
‘ALL PASS’ in the bar at the bottom of the
screen will change to ‘FAILURE’.
• If set to
Off, the Maximum Inactive Channel
power value will be displayed in blue. The bar
at the bottom of the screen will indicate
measurement in progress.
3 Set the high MAC limit.
a. [High MAC Limit]
b. Enter the number on
the numeric keypad.
c. [Enter]
High Limit range: –200 to 200 dB
4 Set the high Pilot limit.
a. [High Pilot Limit]
b. Enter the number on
the numeric keypad.
c. [Enter]
High Limit range: –200 to 200 dB