Technical data

134 Chapter 4
Two Port Insertion Loss
Measuring Two Port Insertion Loss
Two Port Insertion Loss
Measuring Insertion Loss for Power Meter (Option 600 - Power
Meter Only)
NOTE This measurement is only available if you have Option 600 - Power Meter, and
requires an 8482A/8481A or N8481A/N8482A with Option CFT power sensor.
Before making a power measurement, you need to store a PM Loss value that will
compensate for the loss of the devices you use to connect to the transmitter.
Step Notes
1 Connect the cable or
device.
Connect the high power attenuator and remaining
60 cm (2 ft) cable (not the cable used for
normalization) between the two 10 dB pads that
were used for normalization.