Technical data

Chapter 4 139
Two Port Insertion Loss
Measuring Two Port Insertion Loss
Two Port Insertion Loss
Optimizing Dynamic Range
When using the default settings, the test set optimizes measurement accuracy at the
expense of dynamic range. There may be times, however, when you prefer to gain
dynamic range and lose some accuracy. For example, when measuring sector
isolation on adjacent antennas, you may want to see a relatively large dynamic
range, and your results will not be impaired significantly by 1 to 2 dB less accuracy.
Optimizing for dynamic range gains you approximately 30 dB of dynamic range.
Step Notes
1[Setup]
2 On [Optimize] select
Range.
Each time you press this softkey, the selected
option changes.