Technical data
Chapter 5 157
Adjacent Channel Power
Using the Adjacent Channel Power Measurement
Adjacent Channel Power
To Change Frequency Limits:
Setting Up the Signal Generator
“Using the RF OUT Loss field” on page 159
You can set up the signal generator while in the Spectrum Analyzer mode from the
Setup menu. For more detailed information on the signal generator setup and
operation, refer to “Using the Signal Generator” on page 402.
CAUTION The maximum power for the RF In (Port 1) and RF Out/SWR (Port 2) ports is
+20 dBm (100 mW). When using the 8481A/8482A or N8481A/N8482A with
Option CFT Power Sensors, the maximum input power applied to the Power Sensor
is +24 dBm (300 mW). When using the 8481D Power Sensor, the maximum input
Step Notes
1. [Setup]
2. [Frequency Limits]
3. [On] or [Off]
4. Set the upper limit.
a. [Frequency High Limit]
b. Enter the number on the numeric
keypad.
c. [GHz] or [MHz] or [kHz] or
[Hz]
5 Set the lower limit.
a. [Frequency Low Limit]
b. Enter the number on the numeric
keypad.
c. [GHz] or [MHz] or [kHz] or
[Hz]