Technical data
204 Chapter 7
CDMA Analyzer
Optional CDMA Tx Analyzer Measurement Settings
CDMA Analyzer
Setting the Measurement Carrier Feedthrough Upper Limit
2 Set the lower limit.
a. [Low Limit]
b. Enter the number on
the numeric keypad.
c. [Enter]
Lower Limit range: 0 to 1
Step Notes
1 On [Carr Feed-
thru Limit] select
On or
Off.
Enables you to toggle the measurement limits
function
On (active) and Off (inactive).
• If set to
On, the displayed carrier feedthrough
metric will change from blue to red (fail) or
green (pass) and indicate either a “P” (pass) or
“F” (fail). If the measured value exceeds the
upper limit, ALL PASS, in the bar at the bottom
of the screen, will change to FAILURE.
• If set to
Off, the time carrier feedthrough value
will be displayed in blue. The bar at the bottom
of the screen will indicate measurement in
progress.
2 Set the upper limit.
a. [High Limit]
b. Enter the number on
the numeric keypad.
c. [dBc]
Upper Limit range: +100 dBc to –150 dBc
Step Notes