Technical data
Chapter 7 205
CDMA Analyzer
Optional CDMA Tx Analyzer Measurement Settings
CDMA Analyzer
Setting the Measurement Noise Floor Upper Limit
Step Notes
1 On [Noise Floor Limits]
select
On or Off.
Enables you to toggle the measurement limit
function On (active) and Off (inactive).
• If set to
On, the displayed noise floor metrics
will change from blue to red (fail) or green
(pass) and indicate either a “P” (pass) or “F”
(fail). If the measured value exceeds the upper
limit, ALL PASS, in the bar at the bottom of the
screen, will change to FAILURE.
• If set to
Off, the time noise floor value will be
displayed in blue. The bar at the bottom of the
screen will indicate measurement in progress.
2 Set the upper limit.
a. [High Limit]
b. Enter the number on
the numeric keypad.
c. [dB]
Upper Limit range: +100 dBc to –150 dBc