Technical data

206 Chapter 7
CDMA Analyzer
Optional CDMA Tx Analyzer Measurement Settings
CDMA Analyzer
Setting the Measurement Pilot Power Limits
Step Notes
1 On [Pilot Power Limits]
select
On or Off.
Enables you to toggle the measurement limits
function On (active) and Off (inactive).
If set to
On, the delta pilot power metrics will
change from blue to red (fail) or green (pass)
and indicate either a “P” (pass) or “F” (fail). If
the measured value exceeds the upper limit or
lower limit, ALL PASS, in the bar at the bottom
of the screen, will change to FAILURE.
If set to
Off, the time pilot power value will be
displayed in blue. If NO limits are set to On, the
pass/fail bar will not be displayed at the bottom
of the screen.
2 Set the upper limit.
a. [High Limit]
b. Enter the number on
the numeric keypad.
c. [dBm]
Delta Limit range: –150 to 100 dBm
3 Set the lower limit.
a. [Low Limit]
b. Enter the number on
the numeric keypad.
c. [dBm]
Delta Limit range: –150 to 100 dBm