Technical data

208 Chapter 7
CDMA Analyzer
Optional CDMA Tx Analyzer Measurement Settings
CDMA Analyzer
Setting the Measurement Delta Sync Power Limits
Step Notes
1On [Δ Sync Pwr Limits]
select
On or Off.
Enables you to toggle the measurement limits
function On (active) and Off (inactive).
If set to
On, the delta sync. power metrics will
change from blue to red (fail) or green (pass)
and indicate either a “P” (pass) or “F” (fail). If
the measured value exceeds the upper limit or
lower limit, the words ‘ALL PASS’, in the bar
at the bottom of the screen, will change to
‘FAILURE’.
If set to
Off, the time delta sync. power value
will be displayed in blue. If NO limits are set to
On, the pass/fail bar will not be displayed at the
bottom of the screen.
2 Set the upper delta limit.
a. [High Limit]
b. Enter the number on
the numeric keypad.
c. [dB]
Delta Limit range: –150 to 100 dB
2 Set the lower delta limit.
a. [Low Limit]
b. Enter the number on
the numeric keypad.
c. [dB]
Delta Limit range: –150 to 100 dB