Technical data

Chapter 8 241
CDMA Over Air
Optional CDMA Over Air Measurement Settings
CDMA Over Air
Pilot Power Limit: Sets the upper and lower limit of the delta page power. This
limit is set in dBm. Any measured value that is above or below the set limit will
result in a failure indication.
Delta Page Power Limits: Sets the upper and lower limit of the delta page
power. The limit is set in dB and is relative to the pilot power. The Delta Page
Power Limit calculates a value relative to Carrier Power. If the calculated delta
value exceeds the maximum or minimum set value, a failure is indicated.
Delta Sync Power Limits: Sets the upper and lower limit of the delta sync
power. The limit is set in dB and is relative to the pilot power. Any measured
Delta Sync Power value above or below the set limit will result in a failure
indication.
Setting the Measurement Frequency Error Limits
Step Notes
1 [Setup]
2 [More 1 of 2]
3 [Limits]
4 On [Freq Error Limits]
select
On or Off.
Enables you to toggle the measurement limits
function On (active) and Off (inactive).
If set to
On, the Frequency Error will change
from blue to red (fail) or green (pass) and
indicate either a “P” (pass) or “F” (fail). If the
measured value exceed either the upper or
lower limit, ALL PASS, in the bar at the bottom
of the screen, will change to FAILURE.
•If set to
Off, the channel freq/number value will
be displayed in blue. The area at the bottom of
the screen that is used to indicate Pass or Fail
will be over-written. The bar at the bottom of
the screen, will indicate measurement in
progress.