Technical data
Chapter 8 245
CDMA Over Air
Optional CDMA Over Air Measurement Settings
CDMA Over Air
Setting the Measurement Noise Floor Upper Limit
Step Notes
1 On [Noise Floor Limits]
select
On or Off.
Enables you to toggle the measurement limit
function On (active) and Off (inactive).
• If set to
On, the displayed noise floor metrics
will change from blue to red (fail) or green
(pass) and indicate either a “P” (pass) or “F”
(fail). If the measured value exceeds the upper
limit, the words ‘ALL PASS’, in the bar at the
bottom of the screen, will change to
‘FAILURE’.
• If set to
Off, the time noise floor value will be
displayed in blue. The bar at the bottom of the
screen will indicate measurement in progress.
2 Set the upper limit.
a. [High Limit]
b. Enter the number on
the numeric keypad.
c. [dB]
Upper Limit range: 60 to –150 dBc