Technical data

246 Chapter 8
CDMA Over Air
Optional CDMA Over Air Measurement Settings
CDMA Over Air
Setting the Measurement Pilot Power Limits
Step Notes
1 On [Pilot Power Limits]
select
On or Off.
Enables you to toggle the measurement limits
function On (active) and Off (inactive).
If set to
On, the delta pilot power metrics will
change from blue to red (fail) or green (pass)
and indicate either a “P” (pass) or “F” (fail). If
the measured value exceeds the upper limit or
lower limit, the words ‘ALL PASS’, in the bar
at the bottom of the screen, will change to
‘FAILURE’.
If set to
Off, the time pilot power value will be
displayed in blue. If NO limits are set to On, the
pass/fail bar will not be displayed at the bottom
of the screen.
2 Set the upper limit.
a. [High Limit]
b. Enter the number on
the numeric keypad.
c. [dBm]
Delta Limit range: –150 to 100 dBm
3 Set the lower limit.
a. [Low Limit]
b. Enter the number on
the numeric keypad.
c. [dBm]
Delta Limit range: –150 to 100 dBm