Technical data

Chapter 9 267
Channel Scanner
Using the Channel Scanner
Channel Scanner
Setting the Analyzer Input Gain
Normally the E7495A/B is set to add the most gain in the receive path to boost
low-level signals. If the signal is high, there is no reason to add gain. When you’re
measuring a strong signal, select low sensitivity to reduce the gain in the receive
path and lower the noise floor. For a low level signal, select High.
The Sensitivity setting is automatically switched to Low whenever the signal level
at the test set’s connector (that is, after any external attenuation has been applied)
rises above –10 dBm. You can use the Noise Correction feature to recover any lost
dynamic range.
NOTE The Sensitivity setting is coupled to the Noise Correction setting. Changing the
Sensitivity setting from
Low to High will cause the Noise Correction setting to
change from
On to Off. Conversely, if the Noise Correction setting is changed from
Off to On, the Sensitivity setting is changed from High to Low.
Setting Noise Correction
You can turn the Noise Correction function On or Off. When set to On, noise
correction will be activated and correct for the noise generated internally by the
analyzer. It allows the you to make accurate measurements on signals close to the
noise floor.
To Set the Measurement Noise Correction:
Step Notes
1. [Setup]
2. [Sensitivity]
3.
Low or High
Step Notes
1 [Setup]
2 On [Noise Correction]
select
On or Off.
Each time you press this softkey, the selected
option changes.