Technical data
310 Chapter 11
E1 Analyzer
Making E1 Measurements
E1 Analyzer
To Preform a Sample Monitoring Procedure
Step Notes
1 [Mode]
2 [Backhaul]
3 [E1 Analyzer]
4 [Get Started/ Test Mode]
5 Setup monitoring.
a. [Monitor]
b. [Monitor Full E1]
Use this test mode to monitor live data
on the E1 line without affecting service.
The simplest hookup for monitoring
does not require the test Set Tx outputs
to be connected. In some cases, you may
connect the Tx outputs to the E1 circuit.
For those cases, the Tx primary and
secondary transmitted signals are
regenerated from the received signal on
the corresponding line.
Choose [Monitor Full E1] to monitor
performance of a full E1 Circuit.
6 Setup auto configuration.
a. [Auto Config]
b. [Start Primary Auto Config]
Causes the test set to analyze the
incoming E1 signal and to align the
setup choices to that signal.
7 [Back] Shifts focus back to the main mode
selection buttons.
8 Setup the Rx Input.
a. [Setup]
b. [Rx Input]
c. [Terminate]
This assumes the test set is connected
into a jack that interrupts the E1 circuit.
Note: If you are connected to a monitor
port, select [Monitor] instead of
[Terminate].
9 Setup line code.
a. [Line Code]
b. [HDB3]
This assumes the circuit is designed for
HDB3.